The use of contrast agents with fast field-cycling magnetic resonance imaging

Abstract
Fast field-cycling (FFC) MRI allows switching of the magnetic field during an imaging scan. FFC-MRI takes advantage of the T(1) dispersion properties of contrast agents to improve contrast, thus enabling more sensitive detection of the agent. A new contrast agent designed specifically for use with FFC was imaged using both a homebuilt FFC-MRI system and a 3 T Philips clinical MRI scanner. T(1) dispersion curves were obtained using a commercial relaxometer which showed large changes in relaxation rate between fields. A model of magnetization behaviour was used to predict optimum evolution times for the maximum T(1) contrast between samples at each field. Images were processed and analysed to create maps of R(1) values using a set of images at each field. The R(1) maps produced at two different fields were then subtracted from each other in order to create a map of ΔR(1) in which pixel values depend on the change in R(1) of the sample between the two fields. The dispersion properties of the agent resulted in higher contrast in a ΔR(1) image compared with a standard T(1)-weighted image.