Non-Severinghaus Potentiometric Dissolved CO2 Sensor with Improved Characteristics
Open Access
- 22 January 2013
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 85 (3), 1332-1336
- https://doi.org/10.1021/ac303534v
Abstract
A new type of carbon dioxide sensor comprising a pH glass electrode measured against a carbonate-selective membrane electrode based on a tweezer type carbonate ionophore is presented here for the first time. No cumbersome liquid junction based reference element is utilized in this measurement. The sensor shows an expected Nernstian divalent response slope to dissolved CO2 over a wide range covering the routine environmental and physiological PCO2 levels. Unlike the conventional Severinghaus CO2 probe for which the response is substantially delayed to up to 10 min due to diffusion of carbon dioxide into the internal compartment, the ion-selective CO2 sensor proposed here shows a response time (t95%) of 5 s. When used together with a traditional reference electrode, the sensor system is confirmed to also monitor sample pH and carbonate along with carbon dioxide. A selectivity analysis suggests that Cl– does not interfere even at high concentrations, allowing one to explore this type of sensor probe for use in seawater or undiluted blood samples. The CO2 probe has been used in an aquarium to monitor the CO2 levels caused by the diurnal cycles caused by the metabolism of the aquatic plants and shows stable and reproducible results.Keywords
This publication has 34 references indexed in Scilit:
- Surface Plasmon Resonance Sensor for Dissolved and Gaseous Carbon DioxideAnalytical Chemistry, 2012
- Characterization and analytical validation of a microcantilever-based sensor for the determination of total carbonate in soil samplesSensors and Actuators B: Chemical, 2008
- Simultaneous spectrophotometric flow-through measurements of pH, carbon dioxide fugacity, and total inorganic carbon in seawaterAnalytica Chimica Acta, 2007
- Development of an optical sensor probe for the detection of dissolved carbon dioxideSensors and Actuators B: Chemical, 2006
- First electrodes for blood Po2 and Pco2 determinationJournal of Applied Physiology, 2004
- Sensing of Carbon Dioxide by a Decrease in Photoinduced Electron Transfer QuenchingAnalytical Biochemistry, 1999
- Micromachined Severinghaus-Type Carbon Dioxide ElectrodeAnalytical Chemistry, 1999
- Enhanced Serum Carbon Dioxide Measurements with a Silicone Rubber-Based Carbonate Ion-Selective Electrode and a High-pH Dilution BufferAnalytical Chemistry, 1996
- Determination of Blood pCO2Clinical Chemistry, 1961
- Electrodes for Blood pO2 and pCO2 DeterminationJournal of Applied Physiology, 1958