Evaluation of four point-of-care meters for rapid determination of blood lactate concentrations in dogs

Abstract
Objective—To determine whether blood lactate values determined in dogs with 4 commercially available point-of-care meters were in agreement with values determined with a critical care laboratory blood analyzer. Design—Prospective study. Animals—50 dogs evaluated for emergency treatment. Procedures—Blood samples were collected at initial evaluation and processed on 4 point-of-care meters and on a critical care laboratory blood analyzer. Results—All 4 point-of-care lactate meters generated measurements that were in agreement with the hospital's critical care analyzer. Values for agreement (bias) between the 4 point-of-care meters and the critical care analyzer were −0.652 (limits of agreement [LA], −1.958 to 0.654]), −0.670 (LA, −2.110 to 0.769), −0.096 (LA, −2.071 to 1.879), and −0.498 (LA, −2.616 to 1.620), respectively. Conclusions and Clinical Relevance—Despite its prognostic and therapeutic relevance, blood lactate measurement in dogs has been hampered by the inability to perform the test in a timely fashion. Results of the present study indicated that several handheld point-of-care lactate meters provided results that were in agreement with a laboratory critical care blood analyzer.