Low-artifact intravascular devices: MR imaging evaluation.

Abstract
Flow-phantom magentic resonance (MR) imaging, with use of both spin-echo (SE) and gradient-echo (GRE) techniques at 1.5 T, was performed on the percutaneous Greenfield (beta-III titanium alloy [TMA wire]), Amplatz (MP32-N alloy), and Simon nitinol filters and TMA wire facsimiles of the bird''s nest, Gunther, new retrievable, and Amplatz vena caval filters. SE imaging allowed detection of thrombi as small as 5 .times. 5 mm trapped within the percutaneous Greenfield, Simon nitinol, and TMA-wire facsimile filters; with the MP32-N Amplatz filter, a larger volume of thrombus (10 .times. 20-mm clots) was necessary for clot detection. GRE imaging allowed detection of intraluminal tilting of the percutaneous Greenfield and facsimile Amplatz (TMA-wire) filters. GRE imaging was useful for demonstrating postfilter turbulence due to clots, which was greatest for the Amplatz filter. Imaging of facsimile vascular devices made of tantalum or TMA wire did not cause the severe "black-hole" MR artifacts typical of the stainless-steel devices. SE and GRE imaging were very useful for determining caval patency in two patients with previously placed Mobin-Uddin filters. Noninvasive MR evaluation of blood vessels in the presence of a variety of low-artifact intravascular devices appears feasible.