Dose verification of single shot gamma knife applications using VIPAR polymer gel and MRI
- 3 March 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (6), 1235-1250
- https://doi.org/10.1088/0031-9155/50/6/013
Abstract
This work describes an experimental procedure with potential to assess the overall accuracy associated with gamma knife clinical applications, from patient imaging and dosimetry planning to patient positioning and dose delivery using the automated positioning system of a Leksell Gamma Knife model C. The VIPAR polymer gel-MRI dosimetry method is employed due to its inherent three-dimensional feature and linear dose response over the range of gamma knife applications. Different polymer gel vials were irradiated with single shot gamma knife treatment plans using each of the four available collimator helmets to deliver a maximum dose of 30 Gy. Percentage relative dose results are presented not only in the form of one-dimensional profiles but also planar isocontours and isosurfaces in three dimensions. Experimental results are compared with corresponding Gammaplan treatment planning system calculations as well as acceptance test radiochromic film measurements. A good agreement, within the experimental uncertainty, is observed between measured and expected dose distributions. This experimental uncertainty is of the order of one imaging pixel in the MRI gel readout session (<1 mm) and allows for the verification of single shot gamma knife applications in terms of acceptance specifications for precision in beam alignment and accuracy. Averaging net R(2) results in the dose plateau of the 4 mm and 18 mm collimator irradiated gel vials, which were MR scanned in the same session, provides a crude estimate of the 4 mm output factor which agrees within errors with the default value of 0.870.Keywords
This publication has 27 references indexed in Scilit:
- Polymer gel dosimetry using a three-dimensional MRI acquisition techniqueMedical Physics, 2002
- Choice of phantom materials for dosimetry of Leksell Gamma Knife unit: A Monte Carlo studyMedical Physics, 2002
- On the accuracy of monomer/polymer gel dosimetry in the proximity of a high-dose-rate192Ir sourcePhysics in Medicine & Biology, 2001
- Artefacts in multi-echoT2imaging for high-precision gel dosimetry: II. Analysis ofB1-field inhomogeneityPhysics in Medicine & Biology, 2000
- High-resolution dose profile studies based on MR Imaging with polymer BANGTM gels in stereotactic radiation techniquesMagnetic Resonance Imaging, 2000
- An investigation of the chemical stability of a monomer/polymer gel dosimeterPhysics in Medicine & Biology, 2000
- Quality assurance for the Leksell gamma unit: Considering magnetic resonance image-distortion and delineation failure in the targeting of the internal auditory canalMedical Physics, 1999
- Monte Carlo calculation of single‐beam dose profiles used in a gamma knife treatment planning systemMedical Physics, 1998
- Experimental procedure for the manufacture and calibration of polyacrylamide gel (PAG) for magnetic resonance imaging (MRI) radiation dosimetryPhysics in Medicine & Biology, 1998
- Potential human error in setting stereotactic coordinates for radiosurgery: Implications for quality assuranceInternational Journal of Radiation Oncology*Biology*Physics, 1993