Bedeutung physikalischer Parameter für die Effektivität der Radiofrequenz-Katheterablation - Significance of Physical Parameters for the Efficacy of Radiofrequency Catheter Ablation

Abstract
Radiofrequency catheter ablation has been shown to be an effective form of treatment of accessory pathways in patients with WPW-syndrome and other supraventricular tachycardias. However, the biophysical parameters so far used in vivo neither correlated with the size of the myocardial lesion nor did they provide any information about contact of the electrode with the myocardial wall. In this study, 104 radiofrequency energy applications were performed on excised pig myocardium in circulating heparinized pig blood, and in blood alone, and root mean square (rms) voltage, current and phase angle were measured using a specially developed device. The calculated effective power and output power differed by only 2-7% measured at the point of maximum current during coagulation. A progressive drop in current following a rise in impedance led to a phase shift of more than 80 degrees with a decrease in effective power to 17% of the output power. Hence, apparent output power was mainly ineffective power. The time-dependent variations of phase angle, impedance and current were found to be useful for distinguishing between the coagulated media. These results show that physical parameters measured during radio-frequency catheter ablation may help to monitor electrode position in the clinical situation and reduce the number of ineffective energy applications.