Reused silicone catheter for clean intermittent catheterization (CIC): is it safe for spinal cord-injured (SCI) men?
- 3 August 2004
- journal article
- Published by Springer Science and Business Media LLC in Spinal Cord
- Vol. 42 (11), 638-642
- https://doi.org/10.1038/sj.sc.3101646
Abstract
Study Design: Study of reusable catheter. Objective: To investigate whether a silicone cathether reused over years for clean intermittent catheterization (CIC) was safe for spinal cord injured (SCI) men. Setting: Maharaj Hospital, Chiang Mai, Thailand. Method: A cross-sectional study was obtained from SCI men who had used CIC with a reusable silicone catheter for more than a year. Demographic data, urological management and urinary tract complications focusing on the radiologic status of the urethra were reviewed and analyzed. In addition, two reused and one new catheters were studied under electron microscope for catheter morphology (surface and lumen) and stiffness. Results: There were 28 SCI men included in this study. The average duration of CIC use was 4.8 years and the average time of usage for each catheter was 3 years (range 1–7 years). In all, 26 men previously used indwelling catheterization (ID) during the acute phase. In all, 23 men performed self-catheterization. Regarding urinary complications, three reported urethral bleeding, five had episodes of pus per urethra, five had epididymitis, four had passing stones, 18 had occasional foul smelly urine, 10 developed fever and cloudy urine during the past year. Of 17 patients who had ultrasonography done, four had pathologic findings in kidney and one had bladder calculi. Demographic data, urinary management and complications did not have significant relation to the abnormality of the urethrogram or urinary tract infection. However, where the frequency of CIC was higher, the abnormality of the urethra was lower (P<0.05). All had serum Cr level 1.3 mg/dl. Electron microscopic findings of reused catheters for 2 years revealed encrustation but no obstruction in the lumens and 20% increase in stiffness. Conclusion: The clinical outcome, especially with regard to urethral abnormalities with this reusable silicone catheter is as good as with a disposable one. However, to reuse urinary catheters, one should consider the increasing risk of infection. For SCI patients in developing countries, CIC with a reusable silicone catheter may be a suitable and safe choice if one cleans and applies it properly to reduce infection. In order to answer the question how long a person in a developing country should use the same silicone catheter, further research should be conducted.Keywords
This publication has 7 references indexed in Scilit:
- INTERMITTENT CATHETERIZATION WITH A PRELUBRICATED CATHETER IN SPINAL CORD INJURED PATIENTS: A PROSPECTIVE RANDOMIZED CROSSOVER STUDYJournal of Urology, 2001
- Evaluation of the use of Urocath-Gel® catheters for intermittent self-catheterization by male patients using conventional catheters for a long timeSpinal Cord, 2000
- Clean Intermittent Catheterization in Spinal Cord Injury Patients: Long-Term Followup of a Hydrophilic Low Friction TechniqueJournal of Urology, 1995
- Risk Factors for Bacteriuria and Clinical Urinary Tract Infection in Patients Treated with Clean Intermittent CatheterizationJournal of Urology, 1993
- Clean Intermittent Self-Catheterization: a 12-Year FollowupJournal of Urology, 1990
- Fate of Patients Started on Clean Intermittent Self-Catheterization Therapy 10 Years AgoJournal of Urology, 1983
- Complications of Permanent Bladder Catheter Drainage in Spinal Cord Injury PatientsJournal of Urology, 1978