Real‐time interactive MRI on a conventional scanner

Abstract
A real-time interactive MRI system capable of localizing coronary arteries and imaging arrhythmic hearts in real-time is described. Non-2DFT acquisition strategies such as spiral-interleaf, spiral-ring, and circular echo-planar imaging provide short scan times on a conventional scanner. Real-time grid-ding reconstruction at 8–20 image is achieved by distributing the reconstruction on general-purpose UNIX workstations. An X-windows application provides interactive control. A six-interleaf spiral sequence is used for cardiac imaging and can acquire six images/s. A sliding window reconstruction achieves display rates of 16–20 images/s. This allows cardiac images to be acquired in real-time, with minimal motion and flow artifacts, and without breath holding or cardiac gating. Abdominal images are acquired at over 2.5 images/s with spiral-ring or circular echo-planar sequences. Reconstruction rates are 8–10 images/s. Rapid localization in the abdomen is demonstrated with the spiral-ring acquisition, whereas peristaltic motion in the small bowel is well visualized using the circular echo-planar sequence.