Commissioning of a novel microCT/RT system for small animal conformal radiotherapy
- 28 May 2009
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (12), 3727-3740
- https://doi.org/10.1088/0031-9155/54/12/008
Abstract
The purpose of this work was to commission a 120 kVp photon beam produced by a micro-computed tomography (microCT) scanner for use in irradiating mice to therapeutic doses. A variable-aperture collimator has been integrated with a microCT scanner to allow the delivery of beams with pseudocircular profiles of arbitrary width between 0.1 and 6.0 cm. The dose rate at the isocenter of the system was measured using ion chamber and gafchromic EBT film as 1.56-2.13 Gy min(-1) at the water surface for field diameters between 0.2 and 6.0 cm. The dose rate decreases approximately 10% per every 5 mm depth in water for field diameters between 0.5 and 1.0 cm. The flatness, symmetry and penumbra of the beam are 3.6%, 1.0% and 0.5 mm, respectively. These parameters are sufficient to accurately conform the radiation dose delivered to target organs on mice. The irradiated field size is affected principally by the divergence of the beam. In general, the beam has appropriate dosimetric characteristics to accurately deliver the dose to organs inside the mice's bodies. Using multiple beams delivered from a variety of angular directions, targets as small as 2 mm may be irradiated while sparing surrounding tissue. This microCT/RT system is a feasible tool to irradiate mice using treatment planning and delivery methods analogous to those applied to humans.Keywords
This publication has 31 references indexed in Scilit:
- Treatment planning for a small animal using Monte Carlo simulationMedical Physics, 2007
- MicroRT—Small animal conformal irradiatorMedical Physics, 2007
- Design and evaluation of a variable aperture collimator for conformal radiotherapy of small animals using a microCT scannerMedical Physics, 2007
- Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurancePhysics in Medicine & Biology, 2007
- SU-FF-T-152: Development of a Variable-Aperture Collimator for Small Animal Radiation TherapyMedical Physics, 2007
- Advances in Image-Guided Radiation TherapyJournal of Clinical Oncology, 2007
- Post-irradiation colouration of Gafchromic EBT radiochromic filmPhysics in Medicine & Biology, 2005
- Weak energy dependence of EBT gafchromic film dose response in the 50kVp–10MVp X-ray rangeApplied Radiation and Isotopes, 2005
- Mechanism of Impairment in Wound Healing of Electron Beam-Irradiated Rat SkinWound Repair and Regeneration, 2005
- AAPM protocol for 40–300 kV x‐ray beam dosimetry in radiotherapy and radiobiologyMedical Physics, 2001