Development and applicability of a quality control phantom for dental cone-beam CT
Open Access
- 31 December 2010
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 12 (4), 245-260
- https://doi.org/10.1120/jacmp.v12i4.3478
Abstract
Cone-beam CT (CBCT) has shown to be a useful imaging modality for various dentomaxillofacial applications. However, optimization and quality control of dental CBCT devices is hampered due to the lack of an appropriate tool for image quality assessment. To investigate the application of different image quality parameters for CBCT, a prototype polymethyl methacrylate (PMMA) cylindrical phantom with inserts for image quality analysis was developed. Applicability and reproducibility of the phantom were assessed using seven CBCT devices with different scanning protocols. Image quality parameters evaluated were: CT number correlation, contrast resolution, image homogeneity and uniformity, point spread function, and metal artifacts. Deviations of repeated measurements were between 0.0% and 3.3%. Correlation coefficients of CBCT voxel values with CT numbers ranged between 0.68 and 1.00. Contrast-to-noise ratio (CNR) values were much lower for hydroxyapatite (0 < CNR < 7.7) than for air and aluminum (5.0 < CNR < 32.8). Noise values ranged between 35 and 419. The uniformity index was between 3.3% and 11.9%. Full width at half maximum (FWHM) measurements varied between 0.43 mm and 1.07 mm. The increase of mean voxel values surrounding metal objects ranged between 6.7% and 43.0%. Results from preliminary analyses of the prototype quality control phantom showed its potential for routine quality assurance on CBCT. Large differences in image quality performance were seen between CBCT devices. Based on the initial evaluations, the phantom can be optimized and validated.Keywords
Funding Information
- European Atomic Energy Community's Seventh Framework Programme (FP7/2007-2011)
- SEDENTEXCT: Safety and Efficacy of a New and Emerging Dental X-ray Modality (212246)
- NIHR Manchester Biomedical Research Centre
This publication has 23 references indexed in Scilit:
- On cone‐beam computed tomography artifacts induced by titanium implantsClinical Oral Implants Research, 2009
- Evaluation of voxel values in mandibular cancellous bone: relationship between cone‐beam computed tomography and multislice helical computed tomographyClinical Oral Implants Research, 2009
- Study of the scan uniformity from an i-CAT cone beam computed tomography dental imaging systemDento maxillo facial radiology, 2008
- What is Cone-Beam CT and How Does it Work?Dental Clinics of North America, 2008
- Effect of object location on the density measurement and Hounsfield conversion in a NewTom 3G cone beam computed tomography unitDento maxillo facial radiology, 2008
- Restructuring CT dosimetry—A realistic strategy for the future Requiem for the pencil chamberMedical Physics, 2006
- Intraoperative cone‐beam CT for guidance of head and neck surgery: Assessment of dose and image quality using a C‐arm prototypeMedical Physics, 2006
- Assessment of bone segmentation quality of cone-beam CT versus multislice spiral CT: a pilot studyOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2006
- Clinical applications of cone-beam computed tomography in dental practice.2006
- New low‐contrast resolution phantoms for computed tomographyMedical Physics, 1999