Assessment of a circular powered stapler for creation of anastomosis in left-sided colorectal surgery: A prospective cohort study
- 1 December 2020
- journal article
- research article
- Published by Wolters Kluwer Health in International Journal of Surgery
- Vol. 84, 140-146
- https://doi.org/10.1016/j.ijsu.2020.11.001
Abstract
Background: Circular staplers perform a critical function for creation of anastomoses in colorectal surgeries. Powered stapling systems allow for reduced force required by surgeons to fire the device and may provide advantages for creating a secure anastomosis. The objective of this study was to evaluate the clinical performance of a novel circular powered stapler in a post-market setting, during left-sided colectomy procedures. Materials and methods: Consecutive subjects underwent left-sided colorectal resections that included anastomosis performed with the ECHELON CIRCULARTM Powered Stapler (ECP). The primary endpoint was the frequency in which a stapler performance issue was observed. Secondary endpoints included evaluation of ease of use of the device via a surgeon satisfaction questionnaire, and monitoring/recording of procedure-related adverse events (AEs). Results: A total of 168 anastomoses were performed with the ECP. Surgical approaches included robotic-assisted (n = 74, 44.0%), laparoscopic (n = 71, 42.3%), open (n = 20, 11.9%), and hand-assisted minimally invasive (n = 3, 1.8%) procedures. There were 22 occurrences of device performance issues in 20 (11.9%) subjects during surgery. No positive intraoperative leak tests were observed, and only 1 issue was related to a procedure-related AE or surgical complication, which was an instance of incomplete surgical donut necessitating re-anastomosis. Postoperative anastomotic leaks were experienced in 4 (2.4%) subjects. Clavien-Dindo classification of all AEs indicated that 92.0% were Grades I or II. Participating surgeons rated the ECP as easier to use compared to previously used manual circular staplers in 85.7% of procedures. Conclusion: The circular powered stapler exhibited few clinically relevant performance issues, an overall favorable safety profile, and ease of use for creation of left-sided colon anastomoses.Keywords
Funding Information
- Ethicon Endo-Surgery
This publication has 24 references indexed in Scilit:
- Influence of multiple stapler firings used for rectal division on colorectal anastomotic leak rateSurgical Endoscopy, 2017
- Clinical risk factors for anastomotic leakage after laparoscopic anterior resection for rectal cancer: a systematic review and meta-analysisSurgical Endoscopy, 2015
- Risk Factors and Consequences of Anastomotic Leak After ColectomyDiseases of the Colon & Rectum, 2015
- Systematic review of preoperative, intraoperative and postoperative risk factors for colorectal anastomotic leaksBritish Journal of Surgery, 2015
- Endoscopic evaluation of clinical colorectal anastomotic leakageJournal of Surgical Research, 2014
- Increased Local Recurrence and Reduced Survival From Colorectal Cancer Following Anastomotic LeakAnnals of Surgery, 2011
- Risk Factors and Oncologic Impact of Anastomotic Leakage after Rectal Cancer SurgeryDiseases of the Colon & Rectum, 2008
- Prognosis After Anastomotic Leakage in Colorectal SurgeryDiseases of the Colon & Rectum, 2005
- Anastomotic leakage after curative anterior resection results in a higher prevalence of local recurrenceBritish Journal of Surgery, 2003
- Comparison of Manually Constructed and Stapled Anastomoses in Colorectal SurgeryAnnals of Surgery, 1995