Three-dimensional CT with a Modified C-Arm Image Intensifier: Feasibility

Abstract
A portable C arm was modified for cone-beam computed tomography (CT). This three-dimensional (3D) CT imaging system facilitated the acquisition of fluoroscopic images during a 190 degrees rotation and computed a 3D data cube (matrix, 256 x 256 x 256; scanning time, 100 seconds) with multiplanar image reformation. The high-contrast resolution, 0.9 line pairs per millimeter, was comparable; the low-contrast resolution, minimal; and the radiation dose, 60%-80% lower, as compared with these parameters at spiral CT. The normal anatomy of small joints could be depicted, and the osteosynthesis screws in the talus were correctly identified.