Efficacy of an automated external defibrillator in the management of out-of-hospital cardiac arrest: validation of the diagnostic algorithm and initial clinical experience in a rural environment.
- 1 April 1986
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 73 (4), 701-709
- https://doi.org/10.1161/01.cir.73.4.701
Abstract
Automatic external defibrillators (AEDs) may have advantages over manual defibrillation in managing prehospital cardiac arrest, particularly in rural communities. We conducted a two-part evaluation of a commercially available AED. We first established the diagnostic accuracy of the AED's rhythm recognition algorithm by challenging it with 205 cardiac arrest rhythms previously recorded from actual patients in the field. The AED demonstrated 100% specificity and 92% sensitivity for ventricular fibrillation (VF) in this nonclinical setting. We then compared the clinical efficacy of AEDs in 18 small communities (study group) with that of manual defibrillation in 18 additional communities (control group) of similar size. Ambulance technicians using manual defibrillators correctly diagnosed VF more frequently than the AEDs (98% vs 83%; p less than .025). Specificity for VF was similar in the two groups (100% for AEDs vs 94% for technicians; p greater than .10). AEDs were able to deliver shocks more quickly than was possible with the manual defibrillators (1.56 vs 2.77 min; p less than .001). The ability of the AEDs to terminate VF was excellent, converting VF in 28 of 29 (97%) patients to some other rhythm compared with only 37 of 53 (70%) patients in the control group (p less than .01). Hospital admission and discharge rates were similar for the two groups. Ten of the 35 (29%) patients managed with AEDs achieved admission and six (17%) were ultimately discharged. In the control group 17 of 53 (33%) patients with VF were admitted and seven (13%) were discharged (p less than .75). AEDs are an effective alternative to manual defibrillation in small communities.This publication has 19 references indexed in Scilit:
- Limitation on effectiveness of rapid defibrillation by emergency medical technicians in a rural settingAnnals of Emergency Medicine, 1984
- SENSITIVITY, ACCURACY, AND SAFETY OF AN AUTOMATIC EXTERNAL DEFIBRILLATORThe Lancet, 1984
- Out-of-hospital cardiac arrest: Factors associated with survivalAnnals of Emergency Medicine, 1984
- Self-adhesive preapplied electrode pads for defibrillation and cardioversionJournal of the American College of Cardiology, 1984
- Prehospital Defibrillation Performed by Emergency Medical Technicians in Rural CommunitiesNew England Journal of Medicine, 1984
- Prehospital cardiac rhythm deterioration in a system providing only basic life supportAnnals of Emergency Medicine, 1983
- Out-of-hospital resuscitation using the “Heart-Aid”, an automated external defibrillator-pacemakerInternational Journal of Cardiology, 1983
- The Costs of a Suburban Paramedic Program in Reducing Deaths Due to Cardiac ArrestMedical Care, 1981
- Treatment of Out-of-Hospital Cardiac Arrests with Rapid Defibrillation by Emergency Medical TechniciansNew England Journal of Medicine, 1980
- Factors in successful resuscitation by paramedicsJournal of the American College of Emergency Physicians, 1977