Neodymium:YAG laser damage on silicone intraocular lenses

Abstract
In this study we examined 17 explanted silicone intraocular lenses (IOLs) (two one-piece, 15 three-piece) that had evidence of neodymium:YAG (Nd:YAG) laser damage. The majority of lenses (47.1%) had been explanted because of persistent pigment deposits and chronic inflammation. Decentration or IOL dislocation had occurred with 29.4% of lenses. All 17 patients had had Nd:YAG laser posterior capsulectomy, anterior surface polishing, or both; treatments ranged from one to 13 times. By both clinical and gross examination, the laser lesions appeared dark. This was misinterpreted by some opthalmologists as pigment deposits and some patients had additional Nd:YAG laser treatment, which increased the severity of the laser damage. The mean depth of laser damage on the IOLs was 143 +/- 113.4 microns. It was between 51 and 175 microns in approximately 61.0% and deeper than 300 microns (maximum 660 microns) in 9.3%. We compared the explanted IOLs with silicone IOLs that were experimentally damaged using standard power levels from 0.3 to 3.0 mJ and different focal areas. The appearance of the experimentally induced laser burns was equivalent to that on the explanted IOLs, especially the dark laser pits. Since the laser lesions may resemble pigment deposits on slitlamp examination, we recommend careful examination of patients prior to planned laser therapy.