Assessment of Accuracy of PET Utilizing a 3-D Phantom to Simulate the Activity Distribution of [18F]Fluorodeoxyglucose Uptake in the Human Brain
Open Access
- 1 March 1991
- journal article
- Published by SAGE Publications in Journal of Cerebral Blood Flow & Metabolism
- Vol. 11 (1_suppl), A17-A25
- https://doi.org/10.1038/jcbfm.1991.32
Abstract
A three-dimensional brain phantom has been developed to simulate the activity distributions found in human brain studies currently employed in positron emission tomography (PET). The phantom has a single contiguous chamber and utilizes thin layers of lucite to provide apparent relative concentrations of 5, 1, and 0 for gray matter, white matter, and CSF structures, respectively. The phantom and an ideal image set were created from the same set of data. Thus, the user has a basis for comparing measured images with an ideal set that allows a quantitative evaluation of errors in PET studies with an activity distribution similar to that found in patients. The phantom was employed in a study of the effect of deadtime and scatter on accuracy in quantitation on a current PET system. Deadtime correction factors were found to be significant (1.1–2.5) at count rates found in clinical studies. Deadtime correction techniques were found to be accurate to within 5%. Scatter in emission and attenuation correction data consistently caused 5–15% errors in quantitation, whereas correction for scatter in both types of data reduced errors in accuracy to <5%.Keywords
This publication has 13 references indexed in Scilit:
- An investigation of scatter in attenuation correction for PETIEEE Transactions on Nuclear Science, 1989
- An evaluation of a two-dimensional array detector for high resolution PETIEEE Transactions on Medical Imaging, 1988
- Investigation of Count Rate and Deadtime Characteristics of a High Resolution PET SystemJournal of Computer Assisted Tomography, 1988
- Evaluation of the Performance Characteristics of the PC 4600 Positron Emission TomographJournal of Computer Assisted Tomography, 1984
- Photon mass attenuation and energy-absorption coefficientsThe International Journal of Applied Radiation and Isotopes, 1982
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1982
- Quantitation in Positron Emission Computed Tomography: 5. Physical–Anatomical EffectsJournal of Computer Assisted Tomography, 1981
- In vivo quantitation of regional myocardial blood flow by positron-emission computed tomography.Circulation, 1981
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1981
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979