Angioplasty of long or tandem coronary artery lesions using a new longer balloon dilatation catheter: A comparative study
- 1 June 1991
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Diagnosis
- Vol. 23 (2), 84-88
- https://doi.org/10.1002/ccd.1810230203
Abstract
To determine whether or not a new 3 cm long angioplasty balloon was more effective in treatment of long (15–25 mm) or tandem (< 25 mm overall length) coronary lesions, 44 consecutive patients with suitable lesions were alternately assigned to treatment with shorter, standard, 2 cm length, or long, 3 cm balloons. Primary success of PTCA was achieved in 95%. Those treated with long balloons required fewer inflations (3.3 ± 1.5 vs. 5.7 ± 2.1, P=0001) and showed fewer moderate or severe intimal dissections of the lesions (4/22 vs. 12/22, p =.028.) The use of 3cm PTCA balloons in long or tandem lesions appears to be efficacious and less likely to cause dissections than shorter, 2 cm length devices.Keywords
This publication has 4 references indexed in Scilit:
- Guidelines for percutaneous transluminal coronary angioplastyJournal of the American College of Cardiology, 1988
- Angiographic and clinical predictors of acute closure after native vessel coronary angioplasty.Circulation, 1988
- In-hospital morbidity and mortality in patients undergoing elective coronary angioplasty.Circulation, 1985
- Does length or eccentricity of coronary stenoses influence the outcome of transluminal dilatation?Circulation, 1983