Aluminium release by coated and uncoated fluid‐warming devices
- 21 February 2019
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 74 (6), 708-713
- https://doi.org/10.1111/anae.14601
Abstract
The use of fluid‐warming systems is recommended for infusion rates > 500 ml.h−1 to avoid peri‐operative hypothermia. Some fluid‐warming devices use disposable aluminium‐heated plates for heat transfer, but there is no protective coating to separate the fluid from the heated aluminium surface. It is unknown if this could promote release of aluminium into infusion fluids. We investigated a coated (Fluido compact) and an uncoated (enFlow) fluid‐warming device using normal saline or balanced electrolyte solution as infusion fluids, pumped through the heated disposables at flow rates of 2, 4 and 8 ml.min−1 for 60 min each. Aluminium concentrations in the fluid samples were analysed using graphite furnace atomic absorption spectrometry. With saline the coated and uncoated devices yielded aluminium concentrations below the level of quantification (< 128 μg.l−1). Similarly, balanced electrolyte solution in the coated device yielded aluminium concentrations < 128 μg.l−1. However, balanced electrolyte solution in the uncoated device yielded aluminium concentrations of up to 6794 (3465–8002 [1868–7421]) μg.l−1. Repeating this last study at a flow rate of 2 ml.min−1 resulted in quite high aluminium concentrations when the uncoated device was not heated (~1000 μg.l−1) and higher concentrations after the device was heated. We conclude that using uncoated aluminium plates in fluid‐warming systems can lead to a risk of administering potentially harmful concentrations of aluminium when balanced crystalloid solutions are used. The mechanism is unclear, but heat is in part involved. Coating for aluminium within medical devices in direct contact with infusion fluids should be recommended.Keywords
This publication has 20 references indexed in Scilit:
- Aluminium in parenteral nutrition: a systematic reviewEuropean Journal of Clinical Nutrition, 2013
- Aluminium exposure from parenteral nutrition in preterm infants and later health outcomes during childhood and adolescenceProceedings Of The Nutrition Society, 2011
- Aluminum Exposure From Parenteral Nutrition in Preterm Infants: Bone Health at 15-Year Follow-upPediatrics, 2009
- Mechanical Reliability of AMS 700 CX Improved by Parylene CoatingThe Journal of Sexual Medicine, 2009
- Failure of a water-bath design intravenous fluid warmerCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2009
- Human Health Risk Assessment for Aluminium, Aluminium Oxide, and Aluminium HydroxideJournal of Toxicology and Environmental Health, Part B, 2007
- Failure of an iv fluid warming deviceCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2007
- A.S.P.E.N. Statement on Aluminum in Parenteral Nutrition SolutionsNutrition in Clinical Practice, 2004
- Aluminum Neurotoxicity in Preterm Infants Receiving Intravenous-Feeding SolutionsNew England Journal of Medicine, 1997
- Inter-subject variability in the metabolism of aluminium following intravenous injection as citrateHuman & Experimental Toxicology, 1995