A Comparison of Cardiac Output Determined by the Fick Procedure and a Direct Method Using the Rotameter
- 1 June 1950
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 1 (6), 1261-1266
- https://doi.org/10.1161/01.cir.1.6.1261
Abstract
Experiments were performed on the anesthetized dog to determine the accuracy of the widely used Fick procedure. Cardiac output values, recorded by an optical rotameter, served as a reliable standard for comparison. Under the conditions of these experiments, comparisons showed excellent agreement. The average variation between the two series of the measurements was ±5 per cent. Twelve of thirteen comparisons agreed within less than ±8 per cent. This small deviation was within the range of error which might have existed on the basis of known technical inaccuracies.Keywords
This publication has 5 references indexed in Scilit:
- Technic for Measuring Cardiac Output Directly by Cannulation of the Pulmonary ArteryExperimental Biology and Medicine, 1950
- THE EFFECTS OF ANOXEMIC ANOXIA ON EXCITABILITY, CONDUCTION AND REFRACTORINESS OF MAMMALIAN CARDIAC MUSCLEAmerican Journal of Physiology-Legacy Content, 1947
- THE CORONARY CIRCULATION IN THE DOGAmerican Journal of Physiology-Legacy Content, 1947
- DETERMINATION OF CARDIAC OUTPUT IN THE DOG BY THE FICK PROCEDUREAmerican Journal of Physiology-Legacy Content, 1945
- AUGMENTATION OF LEFT CORONARY INFLOW WITH ELEVATION OF LEFT VENTRICULAR PRESSURE AND OBSERVATIONS ON THE MECHANISM FOR INCREASED CORONARY INFLOW WITH INCREASED CARDIAC LOADAmerican Journal of Physiology-Legacy Content, 1944