Adaptive matched filtering for QRS detection
- 1 January 1988
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 147-148 vol.1
- https://doi.org/10.1109/iembs.1988.94450
Abstract
The authors have investigated the use of an adaptive filter to whiten the noise in the ECG (electrocardiogram) signal and adjust the matched filter response accordingly. They applied a simple QRS detection strategy to the filtered signal and evaluated the QRS detector with ECG data containing severe motion artifact and muscle noise. Preliminary results indicate that in the presence of severe motion artifact, adaptive matched filtering for QRS detection represents a significant improvement over application of a matched filter with a white noise assumption.Keywords
This publication has 3 references indexed in Scilit:
- Near-Optimal Signal Preprocessor for Positive Cardiac Arrhythmia IdentificationIEEE Transactions on Biomedical Engineering, 1985
- Adaptive QRS Detection Based on Maximum A Posteriori EstimationIEEE Transactions on Biomedical Engineering, 1982
- Comparison of the convergence of two algorithms for adaptive FIR digital filtersIEEE Transactions on Acoustics, Speech, and Signal Processing, 1981