Abstract
The atrial high rate episode diagnostic in the Thera pacemaker reports frequency, duration, and date/time of atrial tachyarrhythmias according to programmed criteria. The aim of the study was to validate the atrial high rate episode diagnostic feature. Episodes of atrial fibrillation recorded by Holter monitoring were compared to episodes detected by the pacemaker. Forty five ambulatory (Holter) recordings were used for evaluation. Thirty of 45 ambulatory (Holter) recordings showed sinus rhythm. On 4 of these 30 ambulatory (Holter) recordings, the Thera detected 12 episodes of atrial tachyarrhythmias as false-positives (sinus rhythm was detected as atrial tachyarrhythmia). The main reason was far-field R and T was oversensing. On 15 of 45 ambulatory (Holter) recordings, 125 episodes of atrial tachyarrhythmias were recorded. Ninety-three of these events also were detected by the pacemaker, while for 32 events the Thera reported sinus rhythm. The main reason was that the episodes were of too short duration. Therefore, the Thera (programmed with detection rate 160 beats/min, detection beats 40, termination beats 10) was unable to detect atrial tachycardias. Software simulation of the diagnostic algorithm under several programming settings using the digitized Holter files demonstrated highly reliable detection of atrial tachyarrhythmias (sensitivity 98%, specificity 100%) when programmed as follows: detection rate 220 beats/min, detection beats 10, termination beats 20. It can be concluded that Thera's high rate episode monitor is a reliable tool for detection of atrial tachyarrhythmias, if programmed as recommended.Keywords
This publication has 7 references indexed in Scilit:
- Initial Experience with Mode Switching in a Dual Sensor, Dual Chamher Pacemaker in Patients with Paroxysmal Atrial TachyarrhythmiasPacing and Clinical Electrophysiology, 1994
- Dual Chamber, Rate Adaptive Pacing in Patients with Paroxysmal Supraventricular Tachyarrhythmias: Protective Measures for Rate ControlPacing and Clinical Electrophysiology, 1993
- Undesirable Mode Switching with a Dual Chamher Rate Responsive PacemakerPacing and Clinical Electrophysiology, 1993
- A New Feature for Control of Inappropriate High Rate Tracking in DDDR PacemakersPacing and Clinical Electrophysiology, 1990
- Circumvention of Maximum Tracking Limitations with a Rate Modulated Dual Chamber PacemakerPacing and Clinical Electrophysiology, 1989
- Tracking of Atrial Flutter During DDD Pacing: Another Form of Pacemaker‐Mediated TachycardiaPacing and Clinical Electrophysiology, 1984
- AV Universal (DDD) Pacing and Atrial FibrillationClinical Progress in Pacing and Electrophysiology, 1983