Calibration methods in respiratory calorimetry
- 1 April 1984
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 56 (4), 1120-1125
- https://doi.org/10.1152/jappl.1984.56.4.1120
Abstract
Methods for calibration of respiratory gas exchange measurement systems are evaluated. A new method is described in which O2 is removed from the subject chamber or gas stream and measured gravimetrically in a small combustion chamber containing steel wool. The “Fe-burner” method was tested by comparison with the N2-injection method and found to be of similar accuracy. The new method is easy to perform, quick, and readily applicable to both open- and closed-circuit respiratory systems. It has a rate of O2 uptake that can be adjusted to be similar to the rate by an experimental subject (10–600 kg). Complete calibration of open- and closed-circuit indirect calorimetric systems can be achieved when the Fe-burner method, involving measurement of O2 uptake, is combined with the gravimetric injection of specific gases (CO2, methane) produced by the subject.This publication has 1 reference indexed in Scilit:
- A Systematic Method for Validation of Gas Exchange MeasurementsAnesthesiology, 1982