Automated screening of blood cultures with the Malthus microbiological growth analyser.
- 1 June 1988
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 41 (6), 671-675
- https://doi.org/10.1136/jcp.41.6.671
Abstract
A total of 3347 blood cultures from patients in all hospital wards were examined on a Malthus microbiological analyser and by a conventional system. There was no significant difference in the total numbers of positive cultures of clinical importance between the two systems(p > 0 .cntdot. 05). Staphylococcus aureus, however, was isolated more often by the conventional method (p < 0 .cntdot.05). Failure of the automatic detection routine limited the potential of the Malthus system for earlier detection of positive cultures. Daily visual examination of Malthus curves and subculture of bottles not promptly attached to the apparatus were necessary to avoid missing some positive cultures. False positive rates were 13% for the Malthus system and 2% for the conventional system. The contamination rate was considerably lower in Malthus system (p < 0 .cntdot. 001). Further development would be necessary for the apparatus to be acceptable for routine screening of blood cultures.This publication has 9 references indexed in Scilit:
- A note on the use of dextran in blood cultures monitored by conductance methodsJournal of Applied Bacteriology, 1985
- Automated detection of micro-organisms in blood cultures by means of the Malthus Microbiological Growth Analyser.Journal of Clinical Pathology, 1984
- Controlled evaluation of the effect of atmosphere of incubation on detection of bacteremia and fungemia in supplemented peptone brothJournal of Clinical Microbiology, 1982
- Rapid Detection of Bacteremia by an Early Subculture TechnicAmerican Journal of Clinical Pathology, 1975
- Rapid Detection of Bacteremia by a Radiometric System: A Clinical EvaluationAmerican Journal of Clinical Pathology, 1974