B1‐insensitive T2 preparation for improved coronary magnetic resonance angiography at 3 T
Open Access
- 14 March 2006
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 55 (4), 858-864
- https://doi.org/10.1002/mrm.20835
Abstract
At 3 T, the effective wavelength of the RF field is comparable to the dimension of the human body, resulting in B1 standing wave effects and extra variations in phase. This effect is accompanied by an increase in B0 field inhomogeneity compared to 1.5 T. This combination results in nonuniform magnetization preparation by the composite MLEV weighted T2 preparation (T2 Prep) sequence used for coronary magnetic resonance angiography (MRA). A new adiabatic refocusing T2 Prep sequence is presented in which the magnetization is tipped into the transverse plane with a hard RF pulse and refocused using a pair of adiabatic fast-passage RF pulses. The isochromats are subsequently returned to the longitudinal axis using a hard RF pulse. Numerical simulations predict an excellent suppression of artifacts originating from B1 inhomogeneity while achieving good contrast enhancement between coronary arteries and surrounding tissue. This was confirmed by an in vivo study, in which coronary MR angiograms were obtained without a T2 Prep, with an MLEV weighted T2 Prep and the proposed adiabatic T2 Prep. Improved quantitative and qualitative coronary MRA image measurement was achieved using the adiabatic T2 Prep at 3 T. Magn Reson Med, 2006. Published 2006 Wiley-Liss, Inc.Keywords
This publication has 34 references indexed in Scilit:
- Projection imaging of the right coronary artery with an intravenous injection of contrast agentMagnetic Resonance in Medicine, 2004
- Improved three-dimensional free-breathing coronary magnetic resonance angiography using gadocoletic acid (B-22956) for intravascular contrast enhancementJournal of Magnetic Resonance Imaging, 2004
- Coronary Angiography with Magnetization‐Prepared T2 ContrastMagnetic Resonance in Medicine, 1995
- Sweep-diagram analysis of selective adiabatic pulsesJournal of Magnetic Resonance (1969), 1989
- Amplitude- and frequency/phase-modulated refocusing pulses that induce plane rotations even in the presence of inhomogeneous B1 fieldsJournal of Magnetic Resonance (1969), 1988
- Iterative carr-purcell trainsJournal of Magnetic Resonance (1969), 1988
- Amplitude- and frequency-modulated pulses to achieve 90° plane rotations with inhomogeneous B1 fieldsJournal of Magnetic Resonance (1969), 1987
- Highly selective π2 and π pulse generationJournal of Magnetic Resonance (1969), 1984
- Broadband heteronuclear decouplingJournal of Magnetic Resonance (1969), 1982
- RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imagingPhysics in Medicine & Biology, 1978