Aerodynamic Size Distribution of Cromolyn Sodium at Ambient and Airway Humidity1–3
- 1 March 1980
- journal article
- research article
- Published by American Thoracic Society in American Review of Respiratory Disease
- Vol. 121 (3), 513-517
- https://doi.org/10.1164/arrd.1980.121.3.513
Abstract
The aerodynamic size distribution of cromolyn sodium was determined at ambient and high (98 %) humidity using the single particle aerodynamic relaxation time analyzer. This instrument measures the aerodynamic size of single, suspended solid particles and liquid droplets in the respirable size range. The measurement accommodates all properties such as density, shape, and surface characteristics that affect aerodynamic behavior. Particle morphology under ambient humidity conditions was studied using scanning electron microscopy. Count median aerodynamic diameter increased from 1.40 to 1.52 μm (p < 0.025), and the mass median aerodynamic diameter increased from 2.31 to 3.02 μm (p < 0.005) when measured at ambient and high humidities. The total mass of 0.99 mg of active ingredient aerosolized at ambient humidity in the size range compatible with pulmonary deposition was approximately 5 % of the total mass of active ingredient packaged in the capsule. The low mass of particle aerosolized in the size range appropriate for pulmonary deposition may explain the low estimated deposition of cromolyn reported by other investigators.This publication has 1 reference indexed in Scilit:
- AERODYNAMIC SIZE DISTRIBUTION OF METERED-DOSE BRONCHODILATOR AEROSOLSPublished by Elsevier ,1978