Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer
Open Access
- 20 September 2012
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 7 (9), e43702
- https://doi.org/10.1371/journal.pone.0043702
Abstract
Capillary blood sampling has been identified as a potentially suitable technique for use in diagnostic testing of the full blood count (FBC) at the point-of-care (POC), for which a recent need has been highlighted. In this study we assess the accuracy of capillary blood counts and evaluate the potential of a miniaturized cytometer developed for POC testing. Differential leukocyte counts in the normal clinical range from fingerprick (capillary) and venous blood samples were measured and compared using a standard hematology analyzer. The accuracy of our novel microfluidic impedance cytometer (MIC) was then tested by comparing same-site measurements to those obtained with the standard analyzer. The concordance between measurements of fingerprick and venous blood samples using the standard hematology analyzer was high, with no clinically relevant differences observed between the mean differential leukocyte counts. Concordance data between the MIC and the standard analyzer on same-site measurements presented significantly lower leukocyte counts determined by the MIC. This systematic undercount was consistent across the measured (normal) concentration range, suggesting that an internal correction factor could be applied. Differential leukocyte counts obtained from fingerprick samples accurately reflect those from venous blood, which confirms the potential of capillary blood sampling for POC testing of the FBC. Furthermore, the MIC device demonstrated here presents a realistic technology for the future development of FBC and related tests for use at the site of patient care.Keywords
This publication has 17 references indexed in Scilit:
- Integrated systems for rapid point of care (PoC) blood cell analysisLab on a Chip, 2011
- Leukocyte analysis and differentiation using high speed microfluidic single cell impedance cytometryLab on a Chip, 2009
- Guidelines for point‐of‐care testing: haematologyBritish Journal of Haematology, 2008
- Digital signal processing methods for impedance microfluidic cytometryMicrofluidics and Nanofluidics, 2008
- ICSH Guideline for worldwide point‐of‐care testing in haematology with special reference to the complete blood countInternational Journal of Laboratory Hematology, 2008
- Use of capillary blood count parameters in adultsVox Sanguinis, 2007
- Finger-prick blood samples can be used interchangeably with venous samples for CD4 cell counting indicating their potential for use in CD4 rapid testsAIDS, 2007
- Bead-based immunoassays using a micro-chip flow cytometerLab on a Chip, 2007
- Spurious counts and spurious results on haematology analysers: a review. Part II: white blood cells, red blood cells, haemoglobin, red cell indices and reticulocytesInternational Journal of Laboratory Hematology, 2007
- Complete Blood Cell Counts in Capillary and Venous Blood of Healthy Term NewbornsActa Haematologica, 2000