The use of caesium iodide mini scintillation counters for dual isotope pulmonary capillary permeability studies

Abstract
A commercially available system of caesium iodide crystal mini-detectors (Oakfield Instruments, Oxon, UK) was modified so that it was suitable for dual isotopic measurement of the plasma protein accumulation index (PPA)-a measure of pulmonary endothelial permeability. Using this modified system the mean PPA .times. 10-3 min-1 .+-. (S.E.M.) recorded in 11 normal subjects (22 lungs) was 0.18 (0.08) and in 6 patients (9 lungs) with the adult respiratory distress syndrome was 2.88 (0.63) (P < 0.02). These values for PPA concur with those found by other groups using larger sodium iodide detectors. We conclude that with simple modification caesium iodide mini-detectors may be used successfully for the measurement of PPA in the intensive care setting.