Comparison of the efficacy of a demand oxygen delivery system with continuous low flow oxygen in subjects with stable COPD and severe oxygen desaturation on walking.
Open Access
- 1 August 1996
- Vol. 51 (8), 831-834
- https://doi.org/10.1136/thx.51.8.831
Abstract
BACKGROUND: Provision of ambulatory oxygen using an intermittent pulsed flow regulated by a demand oxygen delivery system (DODS) greatly increases the limited supply time of standard portable gaseous cylinders. The efficacy of such a system has not previously been studied during submaximal exercise in subjects with severe chronic obstructive pulmonary disease (COPD) in whom desaturation is likely to be great and where usage is often most appropriate. METHODS: Fifteen subjects with severe COPD and oxygen desaturation underwent six minute walk tests performed in random order to compare the efficacy of a demand oxygen delivery system (DODS) with continuous flow oxygen. Walk distance, breathlessness, oxygen saturation, resting time, and recovery time (objective and subjective) were recorded and compared for each walk. RESULTS: Breathing continuous oxygen compared with baseline air breathing improved mean walk distance (295 m versus 271 m) and recovery time (47 seconds versus 112 seconds), whilst the lowest recorded saturation (81% versus 74%) and time desaturated below 90% (201 seconds versus 299 seconds) were reduced. When the DODS was compared with air breathing only the walk distance changed (283 m versus 271 m). A comparison of the DODS with continuous oxygen breathing showed the DODS to be less effective at oxygenating subjects with inferior lowest saturation (78% versus 81%), time spent below 90% (284 seconds versus 201 seconds), time to objective recovery (83 seconds versus 47 seconds), and walk distance (283 m versus 295 m). CONCLUSIONS: Neither of the delivery systems was able to prevent desaturation in these subjects. The use of continuous flow oxygen, however, was accompanied by improvements in oxygenation, walk distance, and recovery time compared with air breathing. The DODS produced only a small increase in walk distance without elevation of oxygen saturation, but was inferior to continuous flow oxygen in most of the measured variables when compared directly.Keywords
This publication has 18 references indexed in Scilit:
- Arterialised earlobe blood gas analysis: an underused technique.Thorax, 1994
- Portable liquid oxygen and exercise ability in severe respiratory disability.Thorax, 1992
- Comparison of Six Oxygen Delivery Systems for COPD Patients at Rest and during ExerciseChest, 1992
- Comparison of liquid and gaseous oxygen for domiciliary portable use.Thorax, 1992
- Portable oxygen therapy: assessment and usageRespiratory Medicine, 1991
- Demand Oxygen Delivery for Patients with Restrictive Lung DiseaseChest, 1989
- Efficacy of a Pulsed Oxygen Delivery Device during Exercise in Patients with Chronic Respiratory DiseaseChest, 1989
- Performance of a Demand Oxygen Saver System during Rest, Exercise, and Sleep in Hypoxemic PatientsChest, 1988
- Demand Oxygen Delivery during ExerciseChest, 1987
- Arterialized ear lobe blood samples for blood gas tensionsRespiratory Medicine, 1976