LaBr3:Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time
Top Cited Papers
- 19 March 2010
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (7), N179-N189
- https://doi.org/10.1088/0031-9155/55/7/n02
Abstract
The use of time-of-flight (TOF) information in positron emission tomography (PET) enables significant improvement in image noise properties and, therefore, lesion detection. Silicon photomultipliers (SiPMs) are solid-state photosensors that have several advantages over photomultiplier tubes (PMTs). SiPMs are small, essentially transparent to 511 keV gamma rays and insensitive to magnetic fields. This enables novel detector designs aimed at e.g. compactness, high resolution, depth-of-interaction (DOI) correction and MRI compatibility. The goal of the present work is to study the timing performance of SiPMs in combination with LaBr(3):Ce(5%), a relatively new scintillator with promising characteristics for TOF-PET. Measurements were performed with two, bare, 3 mm x 3 mm x 5 mm LaBr(3):Ce(5%) crystals, each coupled to a 3 mm x 3 mm SiPM. Using a (22)Na point source placed at various positions in between the two detectors, a coincidence resolving time (CRT) of approximately 100 ps FWHM for 511 keV annihilation photon pairs was achieved, corresponding to a TOF positioning resolution of approximately 15 mm FWHM. At the same time, pulse height spectra with well-resolved full-energy peaks were obtained. To our knowledge this is the best CRT reported for SiPM-based scintillation detectors to date. It is concluded that SiPM-based scintillation detectors can provide timing resolutions at least as good as detectors based on PMTs.Keywords
This publication has 38 references indexed in Scilit:
- Performance evaluation of SiPM photodetectors for PET imaging in the presence of magnetic fieldsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- The imaging performance of a LaBr3-based PET scannerPhysics in Medicine & Biology, 2009
- After-pulsing and cross-talk in multi-pixel photon countersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2008
- Advances in Yield Calibration of ScintillatorsIEEE Transactions on Nuclear Science, 2008
- Effect of Proton Dose, Crystal Size, and Cerium Concentration on Scintillation Yield and Energy Resolution of LaBr$_{3}$ :CeIEEE Transactions on Nuclear Science, 2007
- Charge carrier and exciton dynamics inscintillators: Experiment and modelPhysical Review B, 2007
- Characterisation of Geiger-mode avalanche photodiodes for medical imaging applicationsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Scintillation detection using 3 mm × 3 mm silicon photomultipliersPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2007
- Limited Geiger-mode microcell silicon photodiode: new resultsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2000
- Avalanche photo diode with local negative feedback sensitive to UV, blue and green lightNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997