Surface modification and evaluation of some commonly used catheter materials. I. Surface properties

Abstract
Double catheter systems consisting of a stiff outer catheter and a flexible, buoyant, flow-directed, inner catheter which is often balloon-tipped have been employed with increasing frequency recently in both therapeutic and diagnostic procedures. Their use, however, has been restricted because of the excessive friction generated between the two catheters. In an attempt to decrease friction between polymers commonly used as catheter materials, oxidation of polyethylene, fluorinated ethylene-propylene copolymer, poly(vinyl chloride), silicone rubber, and polystyrene surfaces was induced by exposing the polymers to radio frequency glow discharge (RFGD) in a helium environment. All polymers were surface characterized utilizing x-ray photoelectron spectroscopy (XPS), scanning electron microscopy (SEM), and contact angle measurements before and after oxidation. This article describes the materials and methods used to fabricate and characterize the polymer surfaces and the results of the characterization. The results indicate that increases in oxygen concentration at the surface of the polymers and decreases in air-water contact angles occur with increased RFGD exposure time. Plateau values were usually obtained after 5–30 s exposure time, yet no apparent changes in surface topography were noted by scanning electron microscopy. The hydrophilic surfaces produced were stable for up to three months storage time in air.