Artifacts in functional magnetic resonance imaging from gaseous oxygen
- 1 July 1995
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 5 (4), 443-445
- https://doi.org/10.1002/jmri.1880050413
Abstract
Unexpectedly large fluctuations in signal intensity wen identified in the functional MRI (FMRI) of normal subjects breathing pure oxygen intermittently. To test the hypothesis that the signal changes were due to fluctuating concentrations of gaseous (paramagnetic) oxygen in the magnetic field, echo planar gradient echo images were acquired of a phantom contiguous to an oxygen mask through which pure oxygen was administered intermittently via plastic tubing. As a control, room air was administered intermittently or oxygen continuously in the same experimental protocol. Signal intensity changes of up to 60% temporally correlated with the administration of oxygen were produced in the phantom. In functional images prepared from the echo planar images, the signal intensity changes resulted in artifacts especially at interfaces in the phantom. The intermittent administration of pure oxygen during acquisition of data for FMRI may produce signal intensity changes that simulate or obscure function.This publication has 4 references indexed in Scilit:
- Echo-planar MR imaging.Radiology, 1994
- Artifacts due to stimulus correlated motion in functional imaging of the brainMagnetic Resonance in Medicine, 1994
- Functional magnetic resonance imaging of complex human movementsNeurology, 1993
- Processing strategies for time‐course data sets in functional mri of the human brainMagnetic Resonance in Medicine, 1993