PreCon: A Fully Automated Interface for the Pre-Gc Concentration of Trace Gases on Air for Isotopic Analysis
- 1 December 1995
- journal article
- research article
- Published by Taylor & Francis in Isotopes in Environmental and Health Studies
- Vol. 31 (3), 277-284
- https://doi.org/10.1080/10256019508036271
Abstract
For the high precision isotope analysis of atmospheric trace gases a computer controlled concentration interface has been developed. From small air samples it collects either N2O or CO2 derived from CH4 at their respective concentrations (0.3 ppm for N2O, 1.7 ppm for CH4) into a small diameter cold trap (−196°C) and interfaces via GC and open split to an isotope ratio mass spectrometer (Finnigan MAT 252) for on-line isotope evaluation. External reproducibilities for repeated measurements of 100 ml air samples from the same source of < 0.2° (δ-notation) have been achieved for 13C/12C from CH4 and for 15N/14N and 18O/16O from N2O. The precision is adequate to monitor the isotopic changes in these gases during a day's course.Keywords
This publication has 3 references indexed in Scilit:
- Acquisition and processing of data for isotope-ratio-monitoring mass spectrometryOrganic Geochemistry, 1994
- Determination of the isotopic composition of atmospheric methane and its application in the AntarcticJournal of Geophysical Research: Atmospheres, 1991
- Isotope-ratio-monitoring gas chromatography-mass spectrometryAnalytical Chemistry, 1978