Abstract
Ferromagnetic metallic implants and materials are regarded as contraindications for MR imaging because of the potential risks associated with their movement or displacement. To date, 14 published articles have evaluated the ferromagnetic qualities of 127 different metallic implants and other materials, including aneurysm and hemostatic clips (32); dental implants and materials (five); intravascular coils, filters, and stents (13); ear implants (14); prosthetic heart valves (29); orthopedic implants and materials (eight); penile implants (nine); and miscellaneous metallic implants and materials (17). All of these materials were evaluated by measuring the deflection forces induced by static magnetic fields at strengths ranging from 0.147 to 4.7 T. This article is a compilation of the results of these studies; it lists all 127 of the materials tested, indicates whether they were found to be deflected by the static magnetic fields, and gives the highest static magnetic field strength at which they were evalua...