Micromethod system for identification of anaerobic bacteria.

  • 1 May 1973
    • journal article
    • Vol. 25 (5), 713-7
Abstract
A micromethod multitest system prepared by Analytab Products, Inc. and conventional tests employed at the Center for Disease Control for identification of anaerobes were compared. All procedures were conducted in an anaerobic glove box. A total of 104 cultures, including 18 reference strains and 86 diagnostic cultures, were examined. Ninety-one percent of the total tests performed with the two systems were in agreement. Greater than 90% agreement between the two systems was obtained with 12 of the 17 differential tests compared. The tests for nitrate reduction and H(2)S production gave the poorest agreement, 77.8 and 80.8%, respectively. Only 66% of the 86 diagnostic cultures could be presumptively identified with the micromethod system supplemented only with microscopy and colonial characteristics. However, when appropriate supplementary tests and gas-liquid chromatography were used with the micromethod system, 85% of the 86 strains could be identified. When Ehrlich reagent, instead of Kovac reagent, was used with the micromethod to test for indole, the agreement in identification was raised to 93%.