Acquiring a four-dimensional computed tomography dataset using an external respiratory signal
Top Cited Papers
- 16 December 2002
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (1), 45-62
- https://doi.org/10.1088/0031-9155/48/1/304
Abstract
Four-dimensional (4D) methods strive to achieve highly conformal radiotherapy, particularly for lung and breast tumours, in the presence of respiratory-induced motion of tumours and normal tissues. Four-dimensional radiotherapy accounts for respiratory motion during imaging, planning and radiation delivery, and requires a 4D CT image in which the internal anatomy motion as a function of the respiratory cycle can be quantified. The aims of our research were (a) to develop a method to acquire 4D CT images from a spiral CT scan using an external respiratory signal and (b) to examine the potential utility of 4D CT imaging. A commercially available respiratory motion monitoring system provided an 'external' tracking signal of the patient's breathing. Simultaneous recording of a TTL 'X-Ray ON' signal from the CT scanner indicated the start time of CT image acquisition, thus facilitating time stamping of all subsequent images. An over-sampled spiral CT scan was acquired using a pitch of 0.5 and scanner rotation time of 1.5 s. Each image from such a scan was sorted into an image bin that corresponded with the phase of the respiratory cycle in which the image was acquired. The complete set of such image bins accumulated over a respiratory cycle constitutes a 4D CT dataset. Four-dimensional CT datasets of a mechanical oscillator phantom and a patient undergoing lung radiotherapy were acquired. Motion artefacts were significantly reduced in the images in the 4D CT dataset compared to the three-dimensional (3D) images, for which respiratory motion was not accounted. Accounting for respiratory motion using 4D CT imaging is feasible and yields images with less distortion than 3D images. 4D images also contain respiratory motion information not available in a 3D CT image.Keywords
This publication has 25 references indexed in Scilit:
- Performance evaluation of an 85-cm-bore X-ray computed tomography scanner designed for radiation oncology and comparison with current diagnostic CT scannersInternational Journal of Radiation Oncology*Biology*Physics, 2002
- Correcting organ motion artifacts in x‐ray CT systems based on tracking of motion phase by the spatial overlap correlator. II. Experimental studyMedical Physics, 2001
- Correcting organ motion artifacts in x‐ray CT medical imaging systems by adaptive processing. I. TheoryMedical Physics, 2001
- Noninvasive Coronary Angiography by Retrospectively ECG-Gated Multislice Spiral CTCirculation, 2000
- A comparison of ventilatory prostate movement in four treatment positionsInternational Journal of Radiation Oncology*Biology*Physics, 2000
- Electrocardiogram‐correlated image reconstruction from subsecond spiral computed tomography scans of the heartMedical Physics, 1998
- Respiration-induced motion of the kidneys in whole abdominal radiotherapy: implications for treatment planning and late toxicityRadiotherapy and Oncology, 1997
- Volumetric transformation of brain anatomyIEEE Transactions on Medical Imaging, 1997
- Respiratory compensation in projection imaging using a magnification and displacement modelIEEE Transactions on Medical Imaging, 1996
- 3D brain mapping using a deformable neuroanatomyPhysics in Medicine & Biology, 1994