Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facility
Top Cited Papers
- 24 June 2004
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 31 (7), 2107-2118
- https://doi.org/10.1118/1.1762792
Abstract
Monte Carlodosimetry calculations are essential methods in radiation therapy. To take full advantage of this tool, the beam delivery system has to be simulated in detail and the initial beam parameters have to be known accurately. The modeling of the beam delivery system itself opens various areas where Monte Carlo calculations prove extremely helpful, such as for design and commissioning of a therapy facility as well as for quality assurance verification. The gantry treatment nozzles at the Northeast Proton Therapy Center (NPTC) at Massachusetts General Hospital (MGH) were modeled in detail using the GEANT4.5.2 Monte Carlo code. For this purpose, various novel solutions for simulating irregular shaped objects in the beam path, like contoured scatterers, patient apertures or patient compensators, were found. The four-dimensional, in time and space, simulation of moving parts, such as the modulator wheel, was implemented. Further, the appropriate physics models and cross sections for proton therapy applications were defined. We present comparisons between measured data and simulations. These show that by modeling the treatment nozzle with millimeter accuracy, it is possible to reproduce measured dose distributions with an accuracy in range and modulation width, in the case of a spread-out Bragg peak (SOBP), of better than 1 mm. The excellent agreement demonstrates that the simulations can even be used to generate beam data for commissioning treatment planning systems. The Monte Carlo nozzle model was used to study mechanical optimization in terms of scattered radiation and secondary radiation in the design of the nozzles. We present simulations on the neutron background. Further, the Monte Carlo calculations supported commissioning efforts in understanding the sensitivity of beam characteristics and how these influence the dosedelivered. We present the sensitivity of dose distributions in water with respect to various beam parameters and geometrical misalignments. This allows the definition of tolerances for quality assurance and the design of quality assurance procedures.Keywords
Funding Information
- National Cancer Institute (CA 21239)
This publication has 27 references indexed in Scilit:
- Biophysical modelling of proton radiation effects based on amorphous track modelsInternational Journal of Radiation Biology, 2001
- Energy Spectra in the NAC Proton Therapy BeamRadiation Protection Dosimetry, 1997
- A general solution to charged particle beam flattening using an optimized dual-scattering-foil technique, with application to proton therapy beamsPhysics in Medicine & Biology, 1994
- Constraints of the multiple‐scattering theory of Molière in Monte Carlo simulations of the transport of charged particlesMedical Physics, 1993
- Stopping powers for the ion-chamber dosimetry of radiotherapeutic heavy-particle beamsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1992
- Compensating for heterogeneities in proton radiation therapyPhysics in Medicine & Biology, 1984
- Energy loss, range, path length, time-of-flight, straggling, multiple scattering, and nuclear interaction probabilityAtomic Data and Nuclear Data Tables, 1982
- Range modulators for protons and heavy ionsNuclear Instruments and Methods, 1975