Reproducibility of coronary lumen, plaque, and vessel wall reconstruction and of endothelial shear stress measurements in vivo in humans
- 12 August 2003
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 60 (1), 67-78
- https://doi.org/10.1002/ccd.10594
Abstract
The purpose of this study was to assess the reproducibility of an in vivo methodology to reconstruct the lumen, plaque, and external elastic membrane (EEM) of coronary arteries and estimate endothelial shear stress (ESS). Ten coronary arteries without significant stenoses (five native and five stented arteries) were investigated. The 3D lumen and EEM boundaries of each coronary artery were determined by fusing end‐diastolic intravascular ultrasound images with biplane coronary angiograms. Coronary flow was measured. Computational fluid dynamics was used to calculate local ESS. Complete data acquisition was then repeated. Analysis was performed on each data set in a blinded manner. The intertest correlation coefficients for all arteries for the two measurements of lumen radius, EEM radius, plaque thickness, and ESS were r = 0.96, 0.96, 0.94, 0.91, respectively (all P values < 0.0001). The 3D anatomy and ESS of human coronary arteries can be reproducibly estimated in vivo. This methodology provides a tool to examine the effect of ESS on atherogenesis, remodeling, and restenosis; the contribution of arterial remodeling and plaque growth to changes in the lumen; and the impact of new therapies. Catheter Cardiovasc Interv 2003;60:67–78. © 2003 Wiley–Liss, Inc.Keywords
This publication has 27 references indexed in Scilit:
- In-vivo prediction of outward remodeling in native portions of stented coronary arteries associated with sites of high endothelial shear stress at the time of deploymentJournal of the American College of Cardiology, 2002
- The influence of the non-Newtonian properties of blood on the flow in large arteries: unsteady flow in a 90° curved tubeJournal of Biomechanics, 1999
- Geometrically correct 3-D reconstruction of intravascular ultrasound images by fusion with biplane angiography-methods and validationIEEE Transactions on Medical Imaging, 1999
- Effects of Curvature and Stenosis-Like Narrowing on Wall Shear Stress in a Coronary Artery Model with Phasic FlowComputers and Biomedical Research, 1997
- Pulsatile Flow in the Human Left Coronary Artery Bifurcation: Average ConditionsJournal of Biomechanical Engineering, 1996
- Shear stress selectively upregulates intercellular adhesion molecule-1 expression in cultured human vascular endothelial cells.Journal of Clinical Investigation, 1994
- Hemodynamics and the Arterial WallJournal of Biomechanical Engineering, 1992
- Effects of Arterial Compliance and Non-Newtonian Rheology on Correlations Between Intimal Thickness and Wall ShearJournal of Biomechanical Engineering, 1992
- Pulsatile non-newtonian blood flow in three-dimensional carotid bifurcation models: a numerical study of flow phenomena under different bifurcation anglesJournal of Biomedical Engineering, 1991
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986