Since August 1984 over 100 samples have been analyzed. 12C, 13C, and 14C isotopes were measured after acceleration and all dates were corrected for natural and preparation-induced fractionation. A precision of better than 1.0% was obtained for samples 10,000 yr BP or less and these results were shown to be reproducible at precision levels as low as 0.3%. An accuracy of better than 1.0% was demonstrated for the same age range by analyzing samples which were previously dated with the conventional technique. The machine background, measured on blank sample holders in a clean ion source, yielded a count rate equivalent of 0.015 ± 0.007% modern.Keywords
This publication has 1 reference indexed in Scilit:
- Isotrace radiocarbon analysis — equipment and proceduresNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1986