The design and test of a new volume coil for high field imaging
- 1 October 1994
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 32 (4), 492-498
- https://doi.org/10.1002/mrm.1910320411
Abstract
A major problem in the development of high field (>100 MHz) large volume (>6000 cm3) MR coils is the interaction of the coil with the subject as well as the radiation loss to the environment. To reduce subject perturbation of the coil resonance modes, a volume coil that uses an array of freely rotating resonant elements radially mounted between two concentric cylinders was designed for operation at 170 MHz. Substantial electromagnetic energy is stored in the resonant elements outside the sample region without compromising the efficiency of the overall coil. This stored energy reduces the effect of the subject on the circuit and maintains a high Q, facilitating the tuning and matching of the coil. The unloaded Q of the coil is 680; when loaded with a head, it was 129. The ratio of 5.3 of the unloaded to loaded Q supports the notion that the efficiency of the coil was maintained in comparison with previous designs. The power requirement and signal-to-noise performance are significantly improved. The coil is tuned by a mechanism that imparts the same degree of rotation on all of the elements simultaneously, varying their degree of mutual coupling and preserving the overall coil symmetry. A thin radiofrequency shield is an integral part of the coil to reduce the radiation effect, which is a significant loss mechanism at high fields. MR images were collected at 4T using this coil design with high sensitivity and B1, homogeneity.Keywords
This publication has 15 references indexed in Scilit:
- Radiative losses of a birdcage resonatorMagnetic Resonance in Medicine, 1993
- Functional mapping of the human visual cortex at 4 and 1.5 tesla using deoxygenation contrast EPIMagnetic Resonance in Medicine, 1993
- Perfusion imagingMagnetic Resonance in Medicine, 1992
- Echo‐planar time course MRI of cat brain oxygenation changesMagnetic Resonance in Medicine, 1991
- In vivo magnetic resonance imaging and spectroscopy of humans with a 4 t whole‐body magnetNMR in Biomedicine, 1990
- Rapid31P spectroscopy on a 4‐T whole‐body systemMagnetic Resonance in Medicine, 1988
- Active magnetic screening of coils for static and time-dependent magnetic field generation in NMR imagingJournal of Physics E: Scientific Instruments, 1986
- An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 TJournal of Magnetic Resonance (1969), 1985
- The sensitivity of the zeugmatographic experiment involving human samplesJournal of Magnetic Resonance (1969), 1979
- The signal-to-noise ratio of the nuclear magnetic resonance experimentJournal of Magnetic Resonance (1969), 1976