Polyethylene Wear Debris and Long-term Clinical Failure of the Charité Disc Prosthesis
- 1 January 2007
- journal article
- case report
- Published by Wolters Kluwer Health in Spine
- Vol. 32 (2), 223-229
- https://doi.org/10.1097/01.brs.0000251370.56327.c6
Abstract
A clinical case series of 4 patients undergoing anterior lumbar revision due to failure of total disc replacement surgery. To assess the clinical significance of polyethylene wear debris in salvage surgery after initial total disc replacement, the pattern and the mechanisms of polyethylene wear in the retrieved cores, and the extent of polyethylene debris in the periprosthetic tissues obtained from 4 patients. Previous in vitro wear tests have demonstrated low wear rates for lumbar artificial discs, suggesting that implant wear may not be a clinically relevant issue with total disc replacement. However, only long-term clinical investigations with analysis of retrieved implants and periprosthetic tissue can ultimately establish the significance of polyethylene wear debris for total disc arthroplasty. Starting in 2004, we began routinely performing salvage procedures in patients with failed total disc replacements. We report on the short-term outcomes of 4 patients at our institution who were revised with a Charité prosthesis (DePuy Spine, Raynham, MA). Wear analysis of the retrieved prosthesis and histologic examination of the periprosthetic tissue were also performed. All of the retrieved polyethylene cores showed evidence of wear, but the extent and severity varied among the 4 patients. Wear and fracture of the core were associated with osteolysis of the underlying sacrum in 1 patient. Histologic examination of the periprosthetic tissues confirmed the presence of wear debris lying in inflammatory fibrous tissue. In 3 of the 4 patients, implant wear was associated with an unfavorable biomechanical environment (e.g., subsidence, migration, undersizing, and adjacent fusion). The mechanisms of wear included adhesive/abrasive wear of the central domed region of the polyethylene core, as well as chronic rim impingement, resulting in rim fatigue and fracture. This study demonstrates the clinical significance of polyethylene wear debris and the potential for osteolysis with total disc replacements. The authors recommend that patients undergoing lumbar disc arthroplasty receive long-term follow-up to monitor the wear and functional status of their implants.Keywords
This publication has 16 references indexed in Scilit:
- Charité total disc replacement—clinical and radiographical results after an average follow-up of 17 yearsEuropean Spine Journal, 2005
- Effects of polyethylene and TiAIV wear particles on expression of RANK, RANKL and OPG mRNAActa Orthopaedica, 2004
- Spinal Implant Debris-Induced OsteolysisSpine, 2003
- Adapting Innovative Motion-Preserving Technology to Spinal Surgical Practice: What Should We Expect to Happen?Spine, 2003
- Basic Science Summary StatementSpine, 2003
- Complications of Artificial Disc ReplacementJournal of Spinal Disorders & Techniques, 2003
- The Biologic Response to Particles From a Lumbar Disc ProsthesisSpine, 2002
- Artificial disc replacement with the modular type SB Charité III: 2-year results in 50 prospectively studied patientsEuropean Spine Journal, 1999
- Results of Disc Prosthesis After a Minimum Follow-Up Period of 2 YearsSpine, 1996
- A Multicenter Retrospective Study of the Clinical Results of the LINK®SB CharitéA Intervertebral Prosthesis. The Initial European ExperienceSpine, 1994