Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice

Abstract
Background— Multislice computed tomography angiography (CTA) is a promising technology for imaging patients with suspected coronary artery disease. Compared with 16-slice CTA, the improved spatial and temporal resolution of 64-slice CTA (0.6- versus 1.0-mm slice thickness and 330- versus 420-ms gantry rotation time) is associated with an increase in radiation dose. The objective of this retrospective investigation was to compare the estimated dose received during 16- and 64-slice CTA in daily practice and to investigate the impact of different scan protocols on dose and image quality. Methods and Results— Radiation dose was estimated for 1035 patients undergoing coronary CTA. Scanning algorithms with and without an ECG-dependent dose modulation and with a reduced tube voltage were investigated on dose estimates and image quality. In the entire patient cohort, radiation dose estimates were 6.4±1.9 and 11.0±4.1 mSv for 16- and 64-slice CTA, respectively (PConclusions— The increase in spatial and temporal resolution with 64-slice CTA is associated with an increased radiation dose for coronary CTA. Dose-saving algorithms are very effective in reducing radiation exposure and should be used whenever possible.