Correction of off resonance‐related distortion in echo‐planar imaging using EPI‐based field maps
- 1 February 1998
- journal article
- other
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 39 (2), 328-330
- https://doi.org/10.1002/mrm.1910390223
Abstract
We present, here, a simple method for measurement and correction of off‐resonance related geometric distortion in echo‐planar imaging (EP1). This method uses high signal‐to‐noise ratio (SNR) EPI‐based field maps, rapidly acquired using a series of gradient recalled images collected across a range of TE values. This field map is distorted in the same manner as the EPI images to be unwarped, providing a direct look‐up table for the correct location of each pixel of data. This method adds very little scan time and is robust and easy to implement.Keywords