The objective of this study was to assess the accuracy and reliability of mercury and aneroid sphygmomanometers. Measurement of accuracy of calibration and evaluation of physical conditions were carried out in 524 sphygmomanometers, 351 from a hospital setting, and 173 from private medical offices. Mercury sphyg- momanometers were considered inaccurate if the meniscus was not ‘0’ at rest. Aneroid sphygmomanometers were tested against a properly calibrated mercury manometer, and were considered calibrated when the error was 3 mm Hg. Both types of sphygmomanometers were evaluated for conditions of cuff/bladder, bulb, pump and valve. Of the mercury sphygmomanometers tested 21% were found to be inaccurate. Of this group, unreliability was noted due to: excessive bouncing (14%), illegibility of the gauge (7%), blockage of the filter (6%), and lack of mercury in the reservoir (3%). Bladder damage was noted in 10% of the hospital devices and in 6% of private medical practices. Rubber aging occurred in 34% and 25%, leaks/holes in 19% and 18%, and leaks in the pump bulb in 16% and 30% of hospital devices and private practice devices, respectively. Of the aneroid sphygmomanometers tested, 44% in the hospital setting and 61% in private medical practices were found to be inaccurate. Of these, the magnitude of inaccuracy was 4–6 mm Hg in 32%, 7–12 mm Hg in 19% and > 13 mm Hg in 7%. In summary, most of the mercury and aneroid sphygmomanometers showed inaccuracy (21% vs 58%) and unreliability (64% vs 70%).