Description of the scatter component in photon-beam data
- 1 January 1988
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 33 (1), 21-32
- https://doi.org/10.1088/0031-9155/33/1/002
Abstract
Kerma scatter-to-primary ratios (SPR) were calculated, using a Monte Carlo code, for water irradiated by photon beams (200 keV monoenergetic, 60Co gamma , 4 and 15 MV X-rays). The SPR at depth d on the centreline of a circular field of radius r can be described by SPR=aZ+bZ2, with Z=rd/(r+d). The proportionality constant a is determined by the linear attenuation coefficient mu and the ratio r/d. The coefficient b for the second-order term depends on r/d and energy, the latter by being proportional to mu 3. This relation between SPR and Z can facilitate analysis of beam data in terms of the contributions from primary and scattered photons, in particular for small depths and small fields where measurements are difficult. This is illustrated by examining measured TAR data for 60Co gamma beams. While the relation between SPR and Z is not fully explained, it is demonstrated qualitatively that SPR approximately=aZ if only first-order scatter is to be accounted for.This publication has 8 references indexed in Scilit:
- Monte Carlo calculations of scatter dose for small field sizes in a 6 0 Co beamMedical Physics, 1986
- Energy and angular distributions of photons from medical linear acceleratorsMedical Physics, 1985
- A note on equivalent circles, squares, and rectanglesMedical Physics, 1982
- Tissue-air ratio and its associated 'in-air' reference minimal phantomPhysics in Medicine & Biology, 1982
- A revised equation relating tissue-air ratio to per cent depth-dosePhysics in Medicine & Biology, 1981
- Brachytherapy build-up factorsThe British Journal of Radiology, 1981
- Output factors and scatter ratios for radiotherapy unitsPhysics in Medicine & Biology, 1978