Abstract
A microscope substage mirror is mounted by means of a channeled lucite block on the base of a Porter-Blum knife holder. The plane face of the mirror, centered on, and about 2 inches below the edge of the knife, reflects light to the front edge of the knife so that the formation of an image of the knife edge on the face of the tissue block results. Alignment of the edge of the knife with the edge of the block to parallelism is accomplished by rotating the block. Alignment of the face of the block to the cutting plane is done by tilting the block until the image of the knife neither approaches the knife nor recedes from it with up and down movement of the block. Ultrathin sections of an area within a 1-2 μ section examined by light microscopy can thus be obtained from a previously established cutting face without loss of material.