Lutetium oxyorthosilicate block detector readout by avalanche photodiode arrays for high resolution animal PET
- 4 September 2004
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 49 (18), 4305-4319
- https://doi.org/10.1088/0031-9155/49/18/008
Abstract
Avalanche photodiodes (APDs) have proven to be useful as light detectors for high resolution positron emission tomography (PET). Their compactness makes these devices excellent candidates for replacing bulky photomultiplier tubes (PMTs) in PET systems where space limitations are an issue. The readout of densely packed, 10 x 10 lutetium oxyorthosilicate (LSO) block detectors (crystal size 2.0 x 2.0 x 12 mm3) with custom-built monolithic 3 x 3 APD arrays was investigated. The APDs had a 5 x 5 mm2 active surface and were arranged on a 6.25 mm pitch. The dead space on the edges of the array was 1.25 mm. The APDs were operated at a bias voltage of approximately 380 V for a gain of 100 and a dark current of 10 nA per APD. The standard deviation in gain between the APDs in the array ranged from 1.8 to 6.5% as the gain was varied from 50 to 108. A fast, low-noise, multi-channel charge sensitive preamplifier application-specific integrated circuit (ASIC) was developed for the APD readout. The amplifier had a rise time of 8 ns, a noise floor of 515 e- rms and a 9 e- pF(-1) noise slope. An acquired flood image showed that all 100 crystals from the block detector could be resolved. Timing measurements with single-channel LSO-APD detectors, as well as with the array, against a plastic scintillator and PMT assembly showed a time resolution of 1.2 ns and 2.5 ns, respectively. The energy resolution measured with a single 4.0 x 4.0 x 10 mm3 LSO crystal, wrapped in four-layer polytetrafluoroethylene (PTFE) tape and coupled with optical grease on a single APD of the array, yielded 15% (full width at half maximum, FWHM) at 511 keV. Stability tests over 9 months of operation showed that the APD arrays do not degrade appreciably. These results demonstrate the ability to decode densely packed LSO scintillation blocks with compact APD arrays. The good timing and energy resolution makes these detectors suitable for high resolution PET.Keywords
This publication has 17 references indexed in Scilit:
- Performance characteristics of a new generation of processing circuits for PET applicationsIEEE Transactions on Nuclear Science, 2003
- Integrated low-noise low-power fast charge-sensitive preamplifier for avalanche photodiodes in JFET-CMOS technologyIEEE Transactions on Nuclear Science, 2001
- A 4/spl times/8 APD array, consisting of two monolithic silicon wafers, coupled to a 32-channel LSO matrix for high-resolution PETIEEE Transactions on Nuclear Science, 2001
- Performance analysis of phoswich/APD detectors and low-noise CMOS preamplifiers for high-resolution PET systemsIEEE Transactions on Nuclear Science, 2001
- A power-efficient, low-noise, wideband, integrated CMOS preamplifier for LSO/APD PET systemsIEEE Transactions on Nuclear Science, 2000
- Optical fiber readout of scintillator arrays using a multi-channel PMT: a high resolution PET detector for animal imagingIEEE Transactions on Nuclear Science, 1996
- Some studies of avalanche photodiode readout of fast scintillatorsIEEE Transactions on Nuclear Science, 1995
- Feasibility study of an avalanche photodiode readout for a high resolution PET with nsec time resolutionIEEE Transactions on Nuclear Science, 1995
- High resolution positron emission tomography with a prototype camera based on solid state scintillation detectorsIEEE Transactions on Nuclear Science, 1990
- A Multicrystal Two Dimensional BGO Detector System for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1986