Knowledge of the interaction forces between surfaces gained using an atomic force microscope (AFM) is crucial in a variety of industrial and scientific applications and necessitates a precise knowledge of the cantilever spring constant. Many methods have been devised to experimentally determine the spring constants of AFM cantilevers. The thermal fluctuation method is elegant but requires a theoretical model of the bending modes. For a rectangular cantilever, this model is available (Butt and Jaschke). Detailed thermal fluctuation measurements of a series of AFM cantilever beams have been performed in order to test the validity and accuracy of the recent theoretical models. The spring constant of rectangular cantilevers can also be determined easily with the method of Sader and White. We found very good agreement between the two methods. In the case of the V-shaped cantilever, we have shown that the thermal fluctuation method is a valid and accurate approach to the evaluation of the spring constant. A comparison between this method and those of Sader-Neumeister and of Ducker has been established. In some cases, we found disagreement between these two methods; the effect of non-conservation of material properties over all cantilevers from a single chip is qualitatively invoked.Keywords
This publication has 25 references indexed in Scilit:
- A Method for the Calibration of Force Microscopy Cantilevers via Hydrodynamic DragLangmuir, 2000
- Dynamic Force Spectroscopy of Molecular Adhesion BondsPhysical Review Letters, 2000
- Elasticity of Single Polymer ChainsMacromolecules, 1999
- How Strong Is a Covalent Bond?Science, 1999
- Determination of the spring constants of probes for force microscopy/spectroscopyNanotechnology, 1996
- Parallel beam approximation for V-shaped atomic force microscope cantileversReview of Scientific Instruments, 1995
- Improvement of thermally induced bending of cantilevers used for atomic force microscopyScanning, 1995
- Calculation of thermal noise in atomic force microscopyNanotechnology, 1995
- Adhesion Forces Between Individual Ligand-Receptor PairsScience, 1994
- Experimental Determination of Spring Constants in Atomic Force MicroscopyLangmuir, 1994