Suppression of tissue harmonics for pulse-inversion contrast imaging using time reversal
Open Access
- 2 September 2008
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 53 (19), 5469-5480
- https://doi.org/10.1088/0031-9155/53/19/013
Abstract
Pulse-inversion (PI) sequences are sensitive to the nonlinear echoes from microbubbles allowing an improvement in the blood-to-tissue contrast. However, at larger mechanical indices, this contrast is reduced by harmonics produced during nonlinear propagation. A method for tissue harmonics cancellation exploiting time reversal is experimentally implemented using a 128-channel 12-bit emitter receiver. The probe calibration is performed by acquiring the nonlinear echo of a wire in water. These distorted pulses are time-reversed, optimized and used for the PI imaging of a tissue phantom. Compared to normal (straight) pulses, the time-reversed distorted pulses reduced the tissue signal in PI by 11 dB. The second harmonic signals from microbubbles flowing in a wall-less vessel were unaffected by the correction. This technique can thus increase the blood-to-tissue contrast ratio while keeping the pressure and the number of pulses constant.Keywords
This publication has 17 references indexed in Scilit:
- Improved contrast ultrasound with tissue harmonic minimizing pulseIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2008
- “Compression-Only” Behavior of Phospholipid-Coated Contrast BubblesJapanese Journal of Clinical Oncology, 2007
- Contrast echocardiography: new agentsJapanese Journal of Clinical Oncology, 2004
- Angiogenesis imaging – ultrasoundThe British Journal of Radiology, 2003
- Improved Detection of Hepatic Metastases with Pulse-Inversion US during the Liver-specific Phase of SHU 508A: Multicenter StudyRadiology, 2003
- Improved contrast to tissue ratio at higher harmonicsUltrasonics, 2002
- Tissue harmonic and contrast-specific imaging: back to gray scale in ultrasoundEuropean Radiology, 2001
- Detection procedures of ultrasound contrast agentsUltrasonics, 2000
- Suppression of propagating second harmonic in ultrasound contrast imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- Distortion and harmonic generation in the nearfield of a finite amplitude sound beamThe Journal of the Acoustical Society of America, 1984