Lens injuries induced by occupational exposure in non-optimized interventional radiology laboratories.
- 1 July 1998
- journal article
- review article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 71 (847), 728-733
- https://doi.org/10.1259/bjr.71.847.9771383
Abstract
Several cases of ophthalmologically confirmed lens injuries, caused by occupational radiation exposure, have occurred in two X-ray rooms devoted to vascular and visceral interventional radiology procedures. Both laboratories were equipped with overcouch X-ray systems not designed for interventional radiology and without specific tools for radiation protection of the eyes. Typical workloads ranged from between two and five procedures per day. For the two radiologists affected, estimates for the dose to eye lens ranged from 450 to 900 mSv per year, over several years. Once the incidents had been detected, the X-ray systems in both rooms were removed and new equipment specifically designed for interventional radiology was installed, including suspended shielding screens. Since these lens injuries were only detected accidentally, measures to avoid similar occurrences in the future are discussed.Keywords
This publication has 21 references indexed in Scilit:
- Staff dose reduction during coronary angiography using low framing speedThe British Journal of Radiology, 1996
- Interventions on the Coronaries: PTCA Versus CABGJournal of Interventional Cardiology, 1996
- Patient and staff dosimetry in neuroradiological proceduresThe British Journal of Radiology, 1995
- The Relationship of Effective Dose to Personnel and Monitor Reading for Simulated Fluoroscopic Irradiation ConditionsHealth Physics, 1993
- Factors affecting the radiation dose to the lens of the eye during cardiac catheterization proceduresThe British Journal of Radiology, 1993
- The dependence of the scattered radiation dose to personnel on technique factors in diagnostic radiologyThe British Journal of Radiology, 1992
- Radiation exposure during radiofrequency catheter ablation of accessory atrioventricular connections.Circulation, 1991
- Dosimetric modelsAnnals of the ICRP, 1989
- An assessment of the radiation dose received by staff using fluoroscopic equipmentThe British Journal of Radiology, 1982