Cleanup of Environmental Sample Extracts Using Florisil Solid-Phase Extraction Cartridges
- 1 May 1989
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Chromatographic Science
- Vol. 27 (5), 209-215
- https://doi.org/10.1093/chromsci/27.5.209
Abstract
Disposable cartridges containing 1 g of Florisil are investigated for cleanup of extracts obtained from various environmental matrices. Elution patterns and recoveries are determined for 22 chlorinated hydrocarbons and 16 phthalate esters in the presence of interferents such as corn oil, diesel hydrocarbons, organochlorine pesticides, and chlorinated phenols. Hexane, hexane/diethyl ether (1:1), hexane/acetone (9:1), and various combinations of hexane/methylene chloride are used as eluants.This publication has 2 references indexed in Scilit:
- Interferences in solid-phase extraction using C-18 bonded porous silica cartridgesAnalytical Chemistry, 1988