Multiple Sample PCR Amplification and Electrophoretic Analysis on a Microchip

Abstract
Polymerase chain reactions (PCRs) were carried out on as many as four DNA samples at a time on a microchip device. The PCR products were then analyzed, either individually or together on the same device, by microchip gel electrophoresis. A standard PCR protocol was used to amplify 199- and 500-base pair (bp) regions of bacteriophage λ DNA and 346- and 410- bp regions of E. coli genomic and plasmid DNAs, respectively. Thermal lysis of the bacteria was integrated into the PCR cycle. A product sizing medium, poly(dimethylacrylamide), and an intercalating dye for fluorescence detection were used in the electrophoretic analysis of the products. PCR product sizes were determined by coelectrophoresis with marker DNA.